SHB_U14 Chapter 8: Calibration
8-27
Flow detector initial calibration, continued
Performing the initial
calibration
Test and calibration software (password required)
Main menu
Calibration
Auxiliary detectors
Start/Span calibration
Selection: Flow detector using á
Enter desired start point: Enter "0" (no flow)
Enter the desired span point value: e.g. set 60 l/h at the sample gas outlet with flow
meter and needle valve
Start calibration with the Start button and follow the menu prompts.
Continued on next page