Messages and tests
100 FSM4000 D184B140U02
8 Messages and tests
8.1 Overview of error states and alarms
State / Error
Flowrate
indicator
display
Notification
with
simulation at
current output
Current
output
Pulse
output
Contact output
Message with
HART
General
Alarm
Min.
alarm
Max.
alarm
Max-Min
alarm
Empty
pipe
Ext.
Diag.
Alarm
0 = "Empty pipe" 0 %
Prog.Al.
DLR
0 % Alarm Alarm More Stat avai.
1 = "AD Converter/DSP" 0 % Prog.Al. 0 % Alarm Trans Mal F.
2 = "Driver" 0 % Prog.Al. 0 % Alarm Trans Mal F.
3= "Flow > 103%" 103 % Yes High.Al 103 % Alarm PV out Limits
4 = "Zero return" 0 % 0 % 0 % Alarm More Stat avai.
5 = "Data base" 0 % Yes Prog.Al. 0 % Alarm Trans Mal F.
6 = "Totalizer" Yes Alarm More Stat avai.
A = "Max Alarm" Yes Alarm Alarm Alarm More Stat avai.
B = "Min Alarm" Yes Alarm Alarm Alarm More Stat avai.
C = "Ext. data base" 0 % Yes Prog.Al. 0 % Alarm Trans Mal F.
F = "FRAM in Primary" 0 % Yes Prog.Al. 0 % Alarm Trans Mal F.
D = "Old Primary" 0 % Yes Prog.Al. 0 % Alarm Trans Mal F.
E = "DC to high" 0 % Yes Prog.Al. 0 % Alarm Trans Mal F.
G = "Error Diagnosis" Alarm Alarm Trans Mal F.