EasyManua.ls Logo

ABB FSM4000-SE41F - Flowmeter Sensor; Flowmeter Sensor SE41 F

ABB FSM4000-SE41F
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Specifications
116 FSM4000 D184B140U02
12.2 Flowmeter sensor
12.2.1 Flowmeter sensor SE41F
Temperature graph
G00214
1
2
3
4
6
5
Fig. 60: Fluid temperature as function of ambient temperature
1 Ambient temperature °C
2 Fluid temperature °C
3 Standard flange (steel): Hard/soft rubber max. 90 / 60 °C
(194 ... 140 °F)
4 Stainless steel flange
5 Standard flange (steel): PTFE / PFA / ETFE max. 130 °C
(266 °F)
6 Hi
g
h temperature: Thick PTFE / PFA max. 180 °C
(
356 °F
)
Max. allowable cleaning temperature PTFE-, PFA-design
CIP cleaning Liner
Flowmeter
sensor
T
max
t
max
Min
T
amb.
Steam cleaning PTFE, PFA 150 °C (302 °F) 60 25 °C (77 °F)
Wet cleaning PTFE, PFA 140 °C (284 °F) 60 25 °C (77 °F)
If the ambient temperature is > 25 °C, then the difference must be
subtracted from the max. cleaning temperature.
T
max
- °C. °C = (T
amb.
- 25 °C)
Weight
See "Dimensions"
Min. permissible pressure as a function of fluid
temperature
Lining Nominal diameter
DN
P
Operation
mbar abs
at
T
Operating
Hard rubber 15 … 1000
(1/2 … 40“)
0 < 90 °C (194 °F)
Soft rubber 50 … 1000
(2 … 40“)
0 < 60 °C (140 °F)
PTFE 10 … 600
(3/8 … 24“)
270
400
500
< 20 °C (68 °F)
< 100 °C (212 °F)
< 130°C (266 °F)
Thick PTFE
high-
temperature
design
25 … 80 (1 … 3“)
100…250 (4…10“)
300 (12“)
0
67
27
< 180 °C (356 °F)
< 180 °C (356 °F)
< 180 °C (356 °F)
PFA 3 … 200
(1/10 … 8“)
0
0
< 130°C (266 °F)
< 180 °C (356 °F)
ETFE 25...1000 (1...40“) 100 < 130 °C (266 °F)
Ceramic
carbide
25 … 1000
(1 … 40“)
0 < 80 °C (176 °F)
Flowmeter sensor material
Parts Standard Others
Lining
PTFE, PFA, hard rubber,
soft rubber, ETFE
Ceramic
carbide
Signal and ground
electrode for
- Hard rubber
- Soft rubber
Stainless steel
1.4571 (316 Ti)
Hastelloy B-3
(2.4600),
Hastelloy C-4
(2.4610),
Titanium,
tantalum,
Platinum-
iridium,
Stainless steel
1.4539 (904 L)
- PTFE, PFA,
ETFE
Hastelloy C-4 (2.4610)
Stainless steel
1.4571 (316 Ti)
Hastelloy B-3
(2.4600),
Titanium,
tantalum,
Platinum-
iridium,
Stainless steel
1.4539 (904 L)
Grounding plate Stainless steel
1.4571 (316 Ti)
On request
Protection plate
Stainless steel
1.4571 (316 Ti)
On request
Process connection material
Parts Standard Others
Flange
DN 3 ... DN 15
(1/10 ... 1/2“)
Stainless steel
1)
Stainless
steel
1)
DN 20 ... DN 300
(3/4 ... 12“)
Galvanized steel
2)
DN 350 ... DN 1000
(14 ... 40“)
Painted steel
2)
Housing
DN 3 ... 300
(1/10 ... 12“)
Dual-shell casing, cast
aluminum, painted, paint
coat, 80 µm thick,
RAL 9002
DN 350 ... DN 1000
(14 ... 40“)
Welded steel design,
painted, paint coat,
80 µm thick, RAL 9002
Terminal box Al alloy, painted,
80 µm dick, frame:
dark gray, RAL 7012
Cover: light gray,
RAL 9002
Meter tube
Stainless steel
1.4301 (304)
cable gland used Polyamide
The process connections are made of one of the materials listed
below:
1) 1.4301 (304), 1.4307, 1.4404 (316L) 1.4435 (316L), 1.4541 (321) 1.4571 (316Ti),
ASTM A182 F304, ASTM A182 F304L, ASTM A182 F316L, ASTM A182 F321, ASTM
A182 F316TI, ASTM A182 F316, 0Cr18Ni9, 0Cr18Ni10, 0Cr17Ni13Mo2,
0Cr27Ni12Mo3, 1Cr18Ni9Ti, 0Cr18Ni12Mo2Ti
2) 1.0038, 1.0460, 1.0570, 1.0432, ASTM A105, Q255A, 20#, 16Mn

Table of Contents