Appendix
D184B140U02 FSM4000 123
13.3 Overview of setting parameters and technical design
Measuring point: TAG no.:
Flowmeter model:
Transmitter type:
Order no.: Device no.: Order no.:
Measured medium temp.: Power supply:
Lining: Electrodes: Exciter frequency:
C
zero
: C
Span
: System zero point:
Parameters Setting range
Prog. Protection code: ................ 0-9999 (0 = factory setting)
Language: ................ e.g., German, English, French, etc.
Flowmeter sensor ................
see name plate or the submenu "System Data / Flowmeter Primary"
Nominal size: ................ DN 1 ... DN 1000
Q
max
: ................ 0.05 Q
max
DN ... 1 Q
max
DN
Pulse factor: ................ Pulse / phys. unit
Pulse width: ................ 0,100 ... 2000 ms
Low cut-off setting: ................ 0 ... 10% of flow range end value
Damping: ................ 0.2 (0.07) ... 20 seconds
Noise reduction ................ OFF / 1 / 2/ 3 / 4
Density: ................ 0.01 g/cm
3
... 5.0 g/cm
3
Unit Q
max
.: ................ e.g., l/s, l/min, l/h, hl/s, hl/min, hl/h, etc.
Unit totalizer: ................ e.g., l, hl, m
3
, igal, gal, etc.
Max. alarm: ................ %
Min. alarm: ................ %
Contact output: ................ Max. alarm, Min. alarm, Max./Min. alarm, General alarm, Empty pipe, etc.
Contact input: ................ External zero return, Totalizer reset, External system zero, no function
Current output: ................ 0/4 ... 20 mA, 0/2 ... 10 mA, 0 ... 5 mA, 0 ... 10-20 mA, 4 ... 12 ... 20 mA
I
out
with alarm: ................ 0 %, 103 %, 3.8 mA, Low, High
Detector e. pipe: ................ ON / OFF
Calibrate e. pipe: ................ 0 ... 10000
Alarm e. pipe: ................ ON / OFF
I
out
with empty pipe: ................ 0 %, 103 %, 3.8 mA, Low, High
Totalizer function: ................ Standard, difference totalizer
Display line 1: ................ Q (%), Q (unit), Q (mA), counter F/R, TAG number, blank line, bar graph
Display line 2: ................ Q (%), Q (unit), Q (mA), counter F/R, TAG number, blank line, bar graph
Display line 3: ................ Q (%), Q (unit), Q (mA), counter F/R, TAG number, blank line, bar graph
Display line 4: ................ Q (%), Q (unit), Q (mA), counter F/R, TAG number, blank line, bar graph
Operating mode: ................ Standard / Piston Pump / Fast
Flow direction: ................ Supply/Return, forward
Directional display: ................ Standard, inverse
Pulse output:
Active Passive
Communication:
HART PROFIBUS PA FOUNDATION Fieldbus Without
Diagnostics
Min-Max threshold
monitoring
Error (E)
Warning (W)
Value E/W Min. Max. Value E/W Min.
Max.
Coil current AC Coil voltage DC
Coil resistance Coil temperature
Coil insulation resistance DAC value
Electrode voltage Electrode balance
Signal-to-noise ratio Magnetic linearity