EasyManua.ls Logo

ABB FSM4000-SE41F - Entering Data in Short Form

ABB FSM4000-SE41F
128 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Parameterization
D184B140U02 FSM4000 61
7.3 Entering data in “short form”
Task Action Display information
Starting point for “Process information”
V98.14 %
V12.000 m
3
1. Open menu.
C/CE
Any parameter
is displayed
2. Search "Prog. Level".
ENTER
"Prog. level"
Disabled
3. Switch off “Programming protection”
Select "Specialist".
STEP or DATA
"Prog. level"
Specialist
4. Set value.
ENTER
"Prog. level"
Specialist
Direct numeric entry Entry from table
Task
Input
Display information Task
Input
Display information
5.
Find parameter
"Qmax".
STEP or
DATA
Qmax
5.
Parameter “Submenu
unit”.
STEP or
DATA
Submenu
1800.00 m³/h
Unit
6. Change parameter
"Qmax".
ENTER Qmax
6. Change parameter
m³/h.
ENTER
Unit Qmax
m³/h
m³/h
Min 500 m³/h
Max 10000 m³/h
7. Find desired value,
e.g., m³/min.
STEP or
DATA
Unit Qmax
m³/min
Qmax
6 2 4 0 , 0 0 m³/h
8. Set value.
ENTER
Unit Qmax
7.
Enter the desired
value in the sequence.
6 x DATA
STEP
2 x DATA
STEP
4 x DATA
STEP
STEP
10 x DATA
STEP
STEP
m³/min
9.
A
new value is
displayed.
Unit Qmax
m³/min
10. Exit menu.
C/CE
8. Set value.
ENTER Qmax
6240.00 m³/h
9.
A
new value is
displayed.
Qmax
6240.00 m³/h
10. Exit menu.
C/CE
11. Search "Prog. Level".
STEP or DATA
"Prog. level"
Specialist
12. Switch on “Programming protection”.
ENTER
"Prog. level"
Specialist
13. Select "Locked"
C/CE
"Prog. level"
Locked

Table of Contents