Safety
D184B140U02 FSM4000 11
1.5.2.3 Factory Tag
The factory tag is located on the flowmeter sensor housing. If the pressure equipment is subject
to the PED (see section 3 para. 3 PED 2014/68/EU), two labels are required:
Pressure equipment subject to PED
S.-Nr.: 0012345
0045
DN50/PN40
Material: 1.4571 / PTFE
Manufactured: 2002 PED: Fluid 1, Gas
ABB Automation Products GmbH
37070 Göttingen - Germany
/ Hast.C-4
1
2
3
4
5
6
G00002
Fig. 4
1 CE mark (with number of labeled
location) to confirm the device meets the
requirements of pressure equipment
directive 2014/68/EU.
2 Serial number for identification of the
pressure equipment by the manufacturer.
3 Nominal size and nominal pressure rating
of pressure equipment.
4 Flange material, liner material and
electrode material (wetted parts).
5 Year of manufacture and specification of
fluid group as per the pressure
equipment directive (PED). Fluid group 1
= hazardous liquids, gaseous.
6 Manufacturer of the pressure equipment.
Pressure equipment outside the applicable range of the PED
S.-Nr.: 0012345
DN50/PN40
Material: 1.4571 / PTFE
Manufactured: 2002 PED: SEP
ABB Automation Products GmbH
37070 Göttingen - Germany
/ Hast.C-4
G00003
Fig. 5
The factory tag contains most of the specifications included on the plate described above with
the following differences:
• There is no CE mark because the pressure equipment, as per section 3 para. 3 of the PED,
is outside the applicable range of the pressure equipment directive 2014/68/EU.
• The reason for the exception is specified in section 3 para. 3 of the PED. The pressure
equipment is categorized as SEP (= sound engineering practice).
Important
If the factory tag is not present, the device is not in compliance with directive 2014/68/EU. The
exception applies for water, power and connected equipment accessories in accordance with