Commissioning
D184B140U02 FSM4000 51
6.3 Easy Set-up: For uncomplicated configuration
The Easy Set-up function enables users to configure the transmitter for quick and convenient
startup. For additional information about configuration options, see the chapter
"Parameterization".
G00563
EASY SETUP
NS
NS
NS
FSM4000
DATA
STEP
ENTER
1
2
3
C/CE
Note!
Parameters can be changed
only when the Program protection
is turned off.
No = STEP
Jump to menu level
Yes = ENTER
Language
English
Unit Qmax
l/s
Unit totalizer
m3
Qmax
50.000 l/s
Damping 1 Tau
0.22 sek
Display mode
4 small
Display 1st line
Q [Unit]
Display 2nd line
Display 4th line
Display 3rd line
Within “Easy Set-up”,
call up the next parameter
using the C/CE key.
Change the parameter
using the
“Enter” function.
[m/s]Q
Pulse width
0.1000 ms
Pulse factor
1/m3
lout High Alarm
23.000 mA
lout Low Alarm
3.6 mA
Iout at Alarm
High Alarm
Totalizer V
Totalizer R
Press any key in the
process display, then use
STEP or DATA to scroll
to the “Easy Set-up“ menu.
“Easy Set-up“ finished.
Jump back to the process display
or use STEP to go to the menu level.
1 Magnet sensor C/CE
2 Magnet sensor
3 Magnet sensor STEP
DATA (> 1s = ENTER)
Fig. 49