Commissioning
48 FSM4000 D184B140U02
The sensor must be installed at a largely vibration-free location.
The sensor and the converter must be assigned properly for the model FSM4000: The
sensors have an end number of X1, X2, etc., on the name plate. The transmitters have the
end numbers Y1, Y2, etc. End numbers X1 and Y1 are considered a unit.
Monitoring of pulse output.
The pulse output can be operated as active output (24 VDC pulse) or as passive output
(optocoupler). The current setting is provided on the name plate for the transmitter.
Modification as shown in the following illustration.
G00559
1
AB
Fig. 48: Setting the pulse output using jumpers
1 Display board A Pulse passive (jumper inner)
B Pulse active
(j
umper outer
)