Parameterization
D184B140U02 FSM4000 67
Submenu/parameter Value range Comment
Span Cs
View only Flowmeter sensor span value Cs.
100.0 %
Zero Cz
View only Flowmeter sensor zero value Cz. For "Older model"
always 0%.
0.0 %
Phase
View only Phase relationship between reference and signal
voltages.
90
Referencevoltage
View only
Display the reference voltage, for "Older Model" selection.
Not displayed for model 10D1422.
70 mV
Meter factor
View only Displayed only if flowmeter sensor 10D1462/72 is
selected
1.0
Order-Number
View only Display the ABB order number for the flowmeter sensor.
ā-----------------------ā
Submenu
Option available only
with "Older Model"
Display required for "Older Model". Use ENTER and
STEP to access the parameters: Mains Frequency, Type
of Sensor, Meter Size and Reference Voltage. Not
displayed, if new sensor (SE41F/SE21) is selected. See
chapter 11.
Primary Setup
Frequency
50, 60 Hz Displays the mains frequency for the flowmeter sensor.
Option available only with new "Flowmeter sensor".
50 Hz
Type of primary
Select the original
model
Display the connected flowmeter sensor model:
DS4_
Meter size
Select the size of the
connected flowmeter
sensor
Select the size of the flowmeter sensor. Option available
only with "Older Model".
100 mm 4 in
Referencevoltage
Enter the reference
voltage
Display the reference voltage, only for "Older Model"
option, see name plate for flowmeter sensor. Not
displayed for models 10D1422, 10D1462/72.
70 mV
Meter factor
Numeric entry
Display the meter factor, only for older model
10D1462/72; see name plate for flowmeter sensor. See
chapter 11.
1.0