FEM630 | ELECTROMAGNETIC FLOWMETER | OI/FEM630-EN REV. A


General Alarm

Only if the parameter ‘Logic Output Action’ is set to Alarm Signal.
Qv Volumeflow Max
Qv Volumeflow Min
EPD
TFE
Gas Bubble
Conductivity
Sensor Temperature
In house Temp
Menu / parameter Description


Logic Output Action
Selection of binary output function.

Active Mode
Select switching properties for the binary output.
Active High: Normally open
Active Low: Normally closed
Default setting: Active High.


Function
Select a function for the digital input.
No function: No function.
Reset all Totalizer: Counter reset for all counters (forward flow, reverse flow and difference totalizer)
Stop all Totalizer: External counter stop for all counters (forward flow, reverse flow and difference
totalizer)

Set Flowrate to zero: Sets flow measurement to 0.
Start/Stop Batching: Start / stop fill operation (only when FillMass function is activated).
Dual Range Mass: Switchover Qm Max 1 / Qm Max 2.
Dual Range Volume: Switchover Qv Max 1 / Qv Max 2.
Active Mode Select switching properties for the digital input.
Delay Time Selection of delay time for suppressing EMC faults on the digital input.
8 ...Operation