ProcessMaster FEP610, HygienicMaster FEH610 | OI/FEP610/FEH610-EN Rev. B  69 
 
Error no. / 
Range 
Error text  Current output  Digital output  Pulse output  LCD display  Error maskable? 
C078.004 / 
Config. 
Flowrate to zero. 
Check digital in terminals. 
4 mA (0 % flow)  Current value - no 
change. 
0 Hz  0 %  Menu "...Group 
Masking". 
C076.006 / 
Config. 
All totalizer stopp. 
Check digital in terminals. 
Current value - no 
change. 
Current value - no 
change. 
Current value - 
no change. 
Current value - 
no change. 
C074.007 / 
Config. 
Totalizer reset. 
Reset of one or more 
Totalizers. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Menu "...Group 
Masking". 
C072.003 / 
Config. 
Simulation is on Simulated 
values. 
Switch off Simulation Mode. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
C070.027 / 
Config. 
An alarm is  simulated. 
Switch off alarm simulation. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
S065.044 / 
Operation 
Inhouse temp. alarm. 
Reduce ambient 
temperature. 
Current value - no 
change. 
No response  No response  Current value - no 
change. 
S064.041 / 
Operation 
EPD alarm. 
Secure pipe is completely 
filled. 
Alarm - as 
configured in 
menu "Curr. at 
EPD Alarm". 
Alarm if DO is 
configured as 
"Logic / Alarm 
Signal / EPD". 
0   0 %  Menu "...Group 
Masking". 
S062.039 / 
Operation 
Sensor temp. limits alarm. 
Change limits or change fluid 
temperature. 
 
Current value - no 
change. 
No response  Current value - no 
change. 
Current value - no 
change. 
Menu "...Group 
Masking". 
S052.017 / 
Operation 
Curr.Out 31 / 32 is 
saturated. 
CO process value out of 
range. Adapt Qmax. 
Alarm - as 
configured in 
menu "Curr.Out > 
20.5mA". 
Current value - no 
change. 
Current value - no 
change. 
Current value - no 
change. 
Menu "...Group 
Masking". 
S047.016 / 
Operation 
Pulse output is cut off. 
Wrong config. Check pulse 
out configuration. 
Current value - no 
change. 
Current value - no 
change. 
Maximum possible 
pulse rate 
Current value - no 
change. 
Menu "...Group 
Masking". 
S046.001 / 
Operation 
Mass flowrate exceeds 
limits. 
Check flowrate and alarm 
limits. 
Current value - no 
change. 
Alarm if DO is 
configured as "Qm 
Massflow Max" or 
"Qm Massflow 
Min". 
Current value - no 
change. 
Current value - no 
change. 
Menu "...Group 
Masking". 
S044.002 / 
Operation 
Volume flowrate exceeds 
limits. 
Check flowrate and alarm 
limits. 
Current value - no 
change. 
Alarm if DO is 
configured as "Qv 
Volumeflow Max" 
or "Qv Volumeflow 
Min". 
Current value - no 
change. 
Current value - no 
change. 
Menu "...Group 
Masking". 
S041.034 / 
Electronics 
DC feedback  regulation. 
Check conditions of 
application. Call service. 
4 mA (0 % flow)  No response.  0 Hz  0 %  Menu "...Group 
Masking". 
S040.031 / 
Electronics 
Coil impedance alarm. 
Call service. 
Current value - no 
change. 
No response.  Current value - no 
change. 
Current value - no 
change. 
Menu "...Group 
Masking".