7.5.3 Menu: Device Info
This menu is only used to display the device parameters. The parameters are displayed independently of the configured access
level, but cannot be changed.
Menu / parameter Description
Device Info
...Sensor Selection of submenu "...Sensor" using .
...Transmitter Selection of submenu "...Transmitter" using .
Device Info / ...Sensor
QmaxDN Type For information purposes only.
Meter Size Nominal size of sensor.
Qm Max DN The value specifies the maximum massflow rate a flow velocity of 10 m/s.
The value is set automatically via the selected nominal size multiplied with the density set.
Qv Max DN The value specifies the maximum volume flowrate rate a flow velocity of 10 m/s.
The value is set automatically via the selected nominal size.
Mains Frequency Mains frequency for the supply power.
Exitation Freq Frequency used to operate the magnet coils of the flowmeter sensor.
Coil Current Regul. For service information purposes only.
Set Point Curr. Current used to operate the magnet coils of the flowmeter sensor.
DC feedback reg. For service information purposes only.
Sensor ID ID number of the sensor.
Sensor Serial No. Serial number of the sensor.
Sensor Run Hours Operating hours of the sensor.
...Se Calibration Selection of submenu "...Se Calibration" using .
Device Info / ...Sensor / ...Se Calibration
Se Span Calibration value in the forward flow and reverse flow direction of the sensor.
Se Zero
First Cal. Date Date of first calibration of sensor (calibration of new device).
Last Cal. Date Date of last calibration of sensor.
Cal. Cert. No. Identification (number) of the relevant calibration certificate.
First Cal. Location Place of first calibration of the sensor.
Last Cal. Location Place of last calibration of sensor.