EasyManuals Logo

Accu-Chek ACTIVE User Manual

Accu-Chek ACTIVE
135 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #108 background imageLoading...
Page #108 background image
108
Symbols, problems and error messages
The meter is turned on, but the display is blank.
The display is defective.
Please contact your customer support and service
centre.
The clock has stopped or the clock is slow.
The meter was exposed to a temperature below
-10 °C and the battery is beginning to freeze.
Turn the meter off. Ensure that the ambient tem-
perature is between +10 and +40 °C and wait for
the meter to adapt to that temperature.
The time is shown as 0:00 or 0:00am and the
date as 0‑ 0.
The meter was exposed to a temperature below
-10 °C and the battery is beginning to freeze.
Turn the meter off. Ensure that the ambient
temperature is between +10 and +40 °C and wait
for the meter to adapt to that temperature.
You have not set the time and date.
Set the time and date (see Chapter 3).

Table of Contents

Other manuals for Accu-Chek ACTIVE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek ACTIVE and is the answer not in the manual?

Accu-Chek ACTIVE Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1 - 2 μL
Test Time5 seconds
Memory Capacity500 results
Display TypeLCD
Automatic CodingYes
Blood SampleCapillary whole blood
Dimensions97.8 x 46.8 x 19.1 mm
Battery Type1 x 3V lithium battery (type CR2032)
Battery LifeApproximately 1000 tests
Data TransferVia infrared (IR) to computer
Operating Conditions8-42°C, 10-85% RH