Testing blood glucose
5.5 Notes on blood glucose testing
Incorrect results can cause the wrong therapy
recommendation to be made and so lead to
serious adverse health effects.
Therefore, please follow the following instructions:
Only use the Accu-Chek Active meter with
Accu-Chek Active test strips that are approved by
Roche Diagnostics GmbH. Other test strips deliver
incorrect results.
If the code number on the display does not match
the code number on the test strip container, do not
perform blood glucose tests. Incorrect coding
produces incorrect results.
Use only test strips with an expiry date which has
not been exceeded. Test strips with an expiry date
that has already passed can produce incorrect
results. If the expiry date has passed, the
symbol will be displayed in the test strip symbol.
Check the expiry date on the label of the test strip
container. You will nd it next to the
symbol.
Also check that the current date is set in the meter
and that the inserted code chip belongs to the test
strip you are using.

Table of Contents

Other manuals for Accu-Chek ACTIVE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek ACTIVE and is the answer not in the manual?

Accu-Chek ACTIVE Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1 - 2 μL
Test Time5 seconds
Memory Capacity500 results
Display TypeLCD
Automatic CodingYes
Blood SampleCapillary whole blood
Dimensions97.8 x 46.8 x 19.1 mm
Battery Type1 x 3V lithium battery (type CR2032)
Battery LifeApproximately 1000 tests
Data TransferVia infrared (IR) to computer
Operating Conditions8-42°C, 10-85% RH