95
Testing blood glucose in more than one patient – Information for healthcare professionals
10.2 Disinfecting the meter
The following parts of the meter may become con-
taminated:
the case
the measuring window cover
the measuring window
The meter, the cover of the measuring window and
the measuring window itself must be thoroughly
cleaned and disinfected at least once a day after use.
Remember to also clean recesses, grooves and gaps.
Cotton swabs, wipes and cloths lightly moistened
with disinfectant are all suitable for disinfecting the
meter.
Use a mixture of 1-propanol, 2-propanol and glutaral-
dehyde, known in some countries under the brand
Bacillol plus.
Allow areas you have wiped time to dry thoroughly.
Make sure that no liquid enters the meter.
Do not spray anything onto the meter and do
not immerse it in liquid. Internal components
could become damaged and thus impair the
proper functioning of the meter.

Table of Contents

Other manuals for Accu-Chek ACTIVE

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Accu-Chek ACTIVE and is the answer not in the manual?

Accu-Chek ACTIVE Specifications

General IconGeneral
Measurement Range10-600 mg/dL
Sample Size1 - 2 μL
Test Time5 seconds
Memory Capacity500 results
Display TypeLCD
Automatic CodingYes
Blood SampleCapillary whole blood
Dimensions97.8 x 46.8 x 19.1 mm
Battery Type1 x 3V lithium battery (type CR2032)
Battery LifeApproximately 1000 tests
Data TransferVia infrared (IR) to computer
Operating Conditions8-42°C, 10-85% RH