EasyManua.ls Logo

Accumet AP75 - Temperature Calibration

Accumet AP75
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
16
Temperature calibration
1. Make sure the cell is attached to the 6-pin
connector. The ATC annunciator will
appear at the right-hand side of the LCD*.
2. Switch the meter on. Press the MODE
key to select conductivity mode.
3. Press the CAL/MEAS key to enter con-
ductivity or TDS calibration mode. The
CAL indicator appears above the primary
display.
See figure
4. While in conductivity or TDS calibration
mode, press the MODE key to enter tem-
perature calibration mode. The primary
display shows the current temperature
reading and the secondary display shows
the factory default temperature value.
See figure
5. Dip the cell into a solution of known
temperature (i.e. a temperature bath).
Allow time for the built-in temperature
sensor to stabilize.
6. Scroll with the and keys to set
the correct temperature value (i.e. the
temperature of the temperature bath).
You can adjust the reading in increments
of 0.1°C.
See figure
7. Once you have selected the correct
temperature, press the ENTER key.
8. Press the CAL/MEAS key to return to
conductivity or TDS measurement mode.
Temperature Calibration
Your probe features a built-in temperature sensor. The temperature sensor is factory
calibrated. Calibrate your sensor only if you suspect temperature errors that may
have occurred over a long period of time or if you have a replacement probe.
4.5
B
B
A
A
Notes
• You can offset the temperature reading up to ±5°C from the original reading.
To exit this program without confirming the temperature calibration value, DO NOT
press ENTER in step 7. Press CAL/MEAS instead.
*If the ATC indicator does not light, see Program P4.3, on page 33 to switch it on.
C
C

Table of Contents

Related product manuals