11. Specifications
Mode Conductivity TDS Temperature
Range 0.00 to 19.99 µS (0.01 µS) 0.00 to 9.99 ppm (0.01 ppm) 0.0 to 100.0°C
(Resolution) 0.0 to 199.9 µS (0.1 µS) 10.0 to 99.9 ppm (0.1 ppm) or
0 to 1999 µS (1 µS) 100 to 999 ppm (1 ppm) 32.0 to 212.0°F
0.00 to 19.99 mS (0.01 mS) 1.00 to 9.99 ppt (0.01 ppt) (0.1°C or °F)
0.0 to 199.9 mS (0.1 mS) 10.0 to 200 ppt (0.1/1 ppt)
Accuracy ±1% full scale ±1% full scale ±0.5°C or °F
Cell constant: 1.0, 0.1, or 10 selectable (cell constant of included cell = 1.0)
Temperature compensation: Automatic or manual 0 to 50°C (32 to 122°F)
Temperature coefficient: 0 to 10% per °C or °F
TDS factor: adjustable from 0.4 to 1.0
Operating temperature: 0 to 50°C (32 to 122°F)
Memory: up to 50 points (includes measurement, temperature, date and time)
Real time clock: stamps calibration data and stored data with time and date (month
and day)
Power: 4 x 1.5 V AAA batteries (>100 hours)
Dimensions: 7.5"L x 3.5"W x 1.75"H (meter only); 9.2"L x 9.2"W x 2.75"H (boxed)
Weight: 1.0 lb (meter only); 2.0 lbs (boxed complete)