EasyManua.ls Logo

Accumet AP75 - 12 Addendum 1: Calibration Tips

Accumet AP75
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
45
12. Addendum 1: Calibration Tips
You only need one calibration for measurement throughout the entire range of the
meter. If a range was not calibrated, the meter automatically detects the closest range
calibrated and uses that calibration information. However, only the ranges that were
calibrated have maximum accuracy.
• If you are measuring in ranges near to or greater than 20 mS (10 ppt), or near to or
lower than 100 µS (50 ppm), calibrate the meter at least once a week to get speci-
fied ±1% F.S. accuracy.
• If you are measuring in the mid ranges and you washed the probe in deionized
water and stored it dry, calibrate the meter at least once a month.
Wet the probe for 10 minutes before calibrating or taking readings to saturate the
probe surface and minimize drift. If you make measurements at extreme tempera-
tures, calibrate the meter at least once a week.
You should only use the conductivity/TDS probe specified for these meters. These
probes have a built-in temperature sensor. If you use a different probe without a
temperature sensor, you must measure the solution temperature separately and
manually enter the solution temperature (see manual temperature compensation on
pages 18-19).

Table of Contents

Related product manuals