16. Addendum 5: Meter factory default settings
Type Parameter Default Remarks
P1.0 Memory clear no Retains current memory
P2.1 Viewing previous – – – No calibration data for 1st range
P2.2
calibration data
– – – No calibration data for 2nd range
P2.3 – – – No calibration data for 3rd range
P2.4 – – – No calibration data for 4th range
P2.5 – – – No calibration data for 5th range
P3.1 Viewing 1.0 No offset for effective cell constant (1st range)
P3.2
probe data
1.0 No offset for effective cell constant (2nd range)
P3.3 1.0 No offset for effective cell constant (3rd range)
P3.4 1.0 No offset for effective cell constant (4th range)
P3.5 1.0 No offset for effective cell constant (5th range)
P4.1 Ready indicator/ READY/ON Ready indicator on; auto endpoint off
auto endpoint
P4.2 Select °C/°F °C Degrees Celsius
P4.3 ATC on or off ATC on
P4.4 TDS factor 0.5 Adjustable from 0.4 to 1.0
P5.1 Temp. coefficient 2.1% per °C Adjustable from 0 to 10%
P5.2 Normalization 25°C Adjustable from 15 to 30°C
temperature
P6.1 Cell constant 1.0 Select from K = 1.0, 0.1 or 10
P8.0 Factory default no Retains your current settings
Notes
The P7.0 clock function retains the year and time that you set, even after reverting to
factory default settings.