EasyManua.ls Logo

Accumet AP75 - Program 6.0: Selecting the Cell Constant

Accumet AP75
52 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
37
P6.0: Selecting the cell constant
Your meter lets you select a cell constant of K = 1.0, 10, or 0.1.
• Use a cell of K = 1.0 for midrange measurements
• Use a cell of K = 10 for high range measurements (above 20 mS or 10 ppt)
• Use a cell of K = 0.1 for low range measurements (below 20 µS or 10 ppm).
The cell included with your meter has a cell constant of K = 1.0. Contact us for infor-
mation on cells with different cell constants.
7.7
From measurement mode
1. Press Setup key to enter Set Up mode.
2. Press the and keys to scroll through
subgroups until you view parameter P6.0.
See figure
3. Press the ENTER key to select
parameter 6.1.
See figure
4. Press the and keys to select the
between K = 1.0, 0.1, or 10.
5. Press the ENTER key to confirm selection
and to return to the subgroup menu. Press
the CAL/MEAS key to return to measure-
ment mode.
Notes
When using a cell of K = 0.1, the lowest
measuring range will be:
• 0 to 1.999 µS (0 to 0.999 ppm).
The 0 to 199.9 mS (10 to 200 ppt) range will
not be accessible.
When using a cell of K = 10, the highest
measuring range will be:
• 0 to 1999 mS (100 to 2000 ppt).
The 0 to 19.99 µS (0.00 to 9.99 ppm) range
will not be accessible.
A
A
B
B

Table of Contents

Related product manuals