18
Manual Temperature Compensation
5.2
From measurement mode
1. Press Setup key to enter Set Up
mode.
2. Press the and keys to scroll
through subgroups until you view
parameter P4.0.
See figure
3. Press the ENTER key three times
to select parameter 4.3. The upper
display shows “ATC” and the lower
display shows “P4.3”.
See figure
4. Press the ENTER key again. The
upper display shows “ATC” and the
lower display shows “YES” or “NO”.
See figure
5. Press the and keys to select
the Automatic Temperature
Compensation feature on of off.
• YES = ATC on
• NO = ATC off
6. Press the ENTER key to confirm
selection and to return to the sub-
group menu. Press the CAL/MEAS
key to return to measurement mode.
IMPORTANT: For manual compensation, you must deactivate the temperature
probe.
5.2.1 Selecting Manual Temperature Compensation
Select between Automatic Temperature Compensation (ATC) and Manual
Temperature Compensation in the Set Up Program P4.3. Meter default is ATC on.
A
A
B
B
C
C