Unpack and Prepare the Navvus® MicroCatheter
34 | ACIST RXi System User’s Guide 901700-001,01 2019-09 English
Basic Operating Procedures
Unpack and prepare the Navvus MicroCatheter by performing these steps:
1. Carefully inspect the Navvus MicroCatheter packaging and components
for damage prior to use.
2. Using sterile technique, open the Navvus MicroCatheter package as
shown:
Technician:
a. Place your le thumb on the plastic corner just under the “Open”
symbol and grip the underside of the tray at the corner with your
ngers.
b. Hold the tray securely with your le hand and use your right hand
to peel the
cover back and o the tray.
c. Present the opened tray to the scrubbed and sterile physician.
Continue to hold the opened tray for the scrubbed and sterile
physician until the product is removed from the tray as described
below.
MPS
Physician/Technician:
d. Grasp the purple handle assembly.
e. Using the same hand and while still holding the purple handle,
grasp the orange coil.
f. Using the same hand and while still holding the purple handle and
orange coil, nally grasp the dispenser coil.
7. Place the dispenser coil and the handle assembly cable in the sterile eld.
Ensure the handle assembly does not get wet.