Megohmmeter Models 5050/5060
2.1 Megohmmeter Features
1. Safetyterminals:“+”,“G”and“-”andaccesstotheprotectivefuse.
2. ACpowerplug(directoperationonACandbatteryrecharge).
3. Back-litliquidcrystaldisplay.
4. RS-232serialinterfacemaleplug(9-pin)
• Forcalibrationonly(Model5050)
• ForconnectiontoaPCorprinter(Model5060)
5. Rotaryselectorswitchwith7positions:
• OFF-Instrumentiso
• MΩ - 500V-Insulationmeasurement(to2TΩ)
• MΩ - 1000V-Insulationmeasurement(to4TΩ)
• MΩ - 2500V-Insulationmeasurement(to10TΩ)
• MΩ - 5000V-Insulationmeasurement(to10TΩ)
• MΩ - 50V-5000V-Insulationmeasurementwithselectabletestvoltage
• SET-UP-Set-upofthemegohmmeter
6. START/STOPbutton
7. 6buttons(5050)or8buttons(5060),eachwitharstandsecondfunction.
Thesecondfunctionsarehighlightedinyellowbeloweachbutton:
• 2
nd
(yellow button)-Selectsthesecondfunctiononeachbutton.
• R-DAR-PI-DD - Lets you choose the type of measurement to be per-
formed. Instantaneous Resistance (R), Dielectric Absorption (DAR),
PolarizationIndex(PI),orDielectricDischarge(DD).
• R(t)-Shows/hidestheintervalvaluesofinsulationresistance,testvolt-
ageandtime-stamping,followingatime-controlledtest.
• V/TIME- Whenmeasuringinsulation,pressingthisbuttondisplaysthe
time elapsed since the beginning of the measurement, then the exact
voltagegenerated.Whenperformingresistanceorcapacitancemeasure-
ment,thisbuttonhasnoeect.InMR(memoryrecall),itdisplaysthedate
andtimeatwhichthemeasurementwasstored,theexacttestvoltage
andtheOBJ:TESTnumber.
•
-Activatesordeactivatesthe“time-controlled”testmode.
• -TurnsthedisplaybacklightONorOFF.
• ALARM-
ActivatesordeactivatesthealarmsprogrammedinSET-UP.
•
-Selectsaparametertobemodied.
• SMOOTH-Stops/startssmoothingofdisplayedvaluesduringinsulation
testing.