Megohmmeter Models 5050/5060
55
8.2 Fuse Replacement
IfFUSE -G-ashesonthedisplay,thefusemustbereplaced,takingalltheneces-
saryprecautionswhenopeninguptheinstrument.
Make sure that none of the terminals are connected and that the selector
switch is set to OFF before opening up the case.
Onlyusethetypeoffuseshownonlabelinsidetheunit’scover:
0.1A-FastActing-380V,5x20mm,10kA
8.3 Cleaning
Disconnect the instrument from any source of electricity.
Useasoftclothlightlydampenedwithsoapywater.Rinsewithawetclothandthen
drywithadrycloth.Donotusealcohol,solventsorhydrocarbons.
8.4 Storage
Iftheinstrumentisnotusedforanextendedperiodoftime,itisrecommendedto
chargetheinstrumenteverytwoorthreemonths.