28
Megohmmeter Models 5050/5060
Capacity Measurement (afterdischargeoftestedcomponent):
Range Resolution Accuracy
0.005 to 9.999µF 1nF
10% of Reading ± 1ct
10.00 to 49.99µF 10nF
Leak Current Measurement
Range Resolution Accuracy
0.000 to 0.250nA
1pA
15% of Reading ± 10cts
0.251 to 9.999nA 10% of Reading
10.00 to 99.99nA 10pA
5% of Reading
100.0 to 999.9nA 100pA
1.000 to 9.999µA 1nA
10.00 to 99.99µA 10nA
100.0 to 999.9µA 100nA
1000 to 3000µA 1µA 10% of Reading
Calculation of Terms DAR and PI
Range Resolution Accuracy
0.02 to 50.00 0.01 5% of Reading ± 1ct
Calculation of Term DD
Range Resolution Accuracy
0.02 to 50.00 0.01 10% of Reading ± 1ct
3.4 Power Supply
• RechargeablebatteryNiMh(8x1.2V/3.5Ah)
• Recharge:85to256V/50-60Hz
Minimum Battery Charge Life(perNFEN61557-2)
Test Voltage Nominal Charge
Number of Measurements
5s on nominal charge (with 25s pause between
each measurement)
500V 500kΩ 6500
1000V 1MΩ 5500
2500V 2.5MΩ 4000
5000V 5MΩ 1500
w ww . . co m