20 1260 Infinity Micro Degasser Service Manual
3 Installing the Micro Vacuum Degasser
Installing the Micro Vacuum Degasser
the degasser to the other modules to shut down the whole system after
an error condition of the degasser.
Figure 4 Rear of the Micro Vacuum Degasser
Figure 5 Front of the Micro Vacuum Degasser
5 Press in the power switch to turn ON the micro vacuum degasser.
NOTE
In an Agilent 1260 Infinity stack, the individual modules are connected through a CAN
cable. The Agilent 1260 Infinity micro vacuum degasser is an exception. The micro vacuum
degasser can be connected via the APG remote cable to the other modules of the stack. An
Agilent 1260 Infinity Instant Pilot can be connected to the CAN bus at any of the modules in
the system except for the degasser. The control software can be connected to the system
through one GPIB or LAN cable (via LAN-Card) at any of the modules (except for the
degasser), preferably at the detector.
EdlZghdX`Zi
GZbdiZXVWaZhdX`Zi
GH'('hdX`Zi
;jhZ]daYZg
HiVijh^cY^XVidg
EdlZghl^iX]
HZg^VacjbWZg
LVhiZdjiaZi
NOTE
The power switch stays pressed in and a green indicator lamp in the power switch is ON
when the micro vacuum degasser is turned ON. When the line power switch stands out and
the green light is OFF, the micro vacuum degasser is turned OFF.