74 1260 Infinity Micro Degasser Service Manual
8 Repairs
Overview of Internal Parts
Exchanging the Vacuum Tubings
1 Remove Top Cover, see “Removing and Refitting the Top Cover” on
page 55.
2 Disconnect and replace the tubings (only one part at a time in order to
avoid confusion about which part should be connected to what
assembly).
3 Disconnect and replace the tubing assembly that connects to the 4
vacuum chambers to the vacuum sensor, proportional valve and vacuum
pump.
4 Disconnect and replace the tubing that connects the two chambers of
the pump.
5 Disconnect and replace the tubing that connects the waste outlet from
the pump out through the front z- panel of the sheet metal kit. Please
refer to the drawing below for correct installation of all tubing. Pay
particular attention to the correct installation of the waste outlet tubing
(drawn in blue color).
When when aged or leaking
Tools required p/n Description
8710-0899 Screwdriver, Pozidriv #1
Parts required # p/n Description
1 PN 5042-8922 Tubing set for G1379B

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals