42 1260 Infinity Micro Degasser Service Manual
5 Optimizing Performance
Increasing the Degasser Performance and Degassing Level
Increasing the Degasser Performance and Degassing Level
For some critical applications it might be useful to increase the
degasser’s performance and degassing level by using two channels of
the degasser in series with only one solvent channel of the pump. For
doing so:
1 Connect the solvent inlet tubing (bottle head assembly, Bottle- head
assembly (G1311- 60003)) coming from the solvent bottle to the inlet
line of the first channel, you want to use.
2 Connect the outlet of the first channel to the inlet if the 2nd solvent
channel of the degasser with the help of the short connecting tubing
(Connecting tubing (to connect to channels in series for increased
performance) 2x (G1379- 68706)), delivered with the accessory kit of the
degasser.
3 Connect a degasser to pump connecting tubing (Kit of 4 solvent tubes
for connection degasser to MCGV including labels (G1322- 67300)) to the
outlet of the 2nd channel of the degasser and into the pump.
Figure 9 Connecting two degasser channels in series
idejbe
[gdbWdiiaZ

Table of Contents

Other manuals for Agilent Technologies 1260

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1260 and is the answer not in the manual?

Agilent Technologies 1260 Specifications

General IconGeneral
BrandAgilent Technologies
Model1260
CategorySecurity Sensors
LanguageEnglish

Related product manuals