EasyManua.ls Logo

Agilent Technologies 1260 - Exchanging the Proportional Valve

Agilent Technologies 1260
108 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
72 1260 Infinity Micro Degasser Service Manual
8 Repairs
Overview of Internal Parts
Exchanging the Proportional Valve
1 Remove Top Cover, see “Removing and Refitting the Top Cover” on
page 55.
2 Disconnect the valve connector from the control board (CN8) (for
identifying internal parts see “Overview of Internal Parts” on page 66).
3 Loosen and remove the holding screws of the valve.
4 Loosen and remove the two screws of the holding plate at the valve.
5 Disconnect inlet and outlet tubes from the valve.
6 Fix the holding plate to the new valve.
7 Connect the two vacuum tubes to the valve. The inlet of the valve is
labeled IN and is connected to the vacuum pump.
8 Fix the valve with the two screws to its holder.
9 Connect the valve cable to the control board (CN8).
When Valve not switching correctly or leaking
Tools required p/n Description
8710-0899 Screwdriver, Pozidriv #1
Parts required # p/n Description
1 G1379-60003 Proportional valve

Table of Contents

Other manuals for Agilent Technologies 1260

Related product manuals