1260 Infinity Micro Degasser Service Manual 11
Site Requirements and Specifications
2
Physical Specifications
Physical Specifications
Tabl e 1 Physical Specifications
Type Specification Comments
Weight 7 kg (16 lbs)
Dimensions (height ×
width × depth)
80 x 345 × 435 mm (3.1 x 13.5 × 17 inches)
Line voltage 100 – 240 VAC, ± 10 % Wide-ranging
capability
Line frequency 50 or 60 Hz, ± 5 %
Power consumption 30 VA / 30 W / 102 BTU Maximum
Ambient operating
temperature
4–45 °C (39–113 °F)
Ambient non-operating
temperature
-40 – 70 °C (-40 – 158 °F)
Humidity < 95 % r.h. at 40 °C (104 °F) Non-condensing
Operating altitude Up to 2000 m (6562 ft)
Non-operating altitude Up to 4600 m (15091 ft) For storing the module
Safety standards: IEC, CSA,
UL
Installation category II, Pollution degree 2 For indoor use only.