26 1260 Infinity Micro Degasser Service Manual
3 Installing the Micro Vacuum Degasser
Operational Hints for the Micro Vacuum Degasser
When priming the micro vacuum degasser with a syringe the solvent is drawn through the
degasser tubes very quickly. The solvent at the degasser outlet will therefore not be fully
degassed. Pump for approximately 5 minutes with your selected flow rate before starting
any application. This will allow the micro vacuum degasser to properly degas the solvent in
the degasser tubes.