1260 Infinity II Analytical-Scale & Bio-inert Fraction Collector User Manual 85
Preparing the Fraction Collector
4
Method Parameter Settings
Table 1 3 Timetable Functions
Function Parameters
Fraction Mode
•Off
(Turns off the current fraction collection, where
you use Off to turn off fraction collection at the end of
the run if you have not specified a
Stoptime
)
• Time-based, collecting a number of fractions
(Fractions are collected between this time and the
next change of fraction mode or Off, where you
specify the number of fractions to collect in the
Number of Fractions
field)
• Time-based, collecting time slices
(Time-slice
fractions are collected between this time and the next
change of fraction mode or Off. where you specify the
duration of the time-slices to collect in the
Time
slices
field)
• Time-based, collecting volume slices
(Volume-slice
fractions are collected between this time and the next
change of fraction mode or Off, where you specify the
volume of the fractions to collect in the
Volume slices
field)
•Peak-based
(Fractions are collected based on the
peak detection settings)
• Peak-based, collecting time slices
(Time-slice
fractions are collected during the elution of a peak,
based on the peak detection settings, where you
specify the duration of the time-slices to collect in the
Time slices
field)
• Peak-based, collecting volume slices
(Volume-slice
fractions are collected during the elution of a peak,
based on the peak detection settings, where you
specify the volume of the fractions to collect in the
Volume slices
field)
• Peak-based with time-slice recovery
(Time-slice
fractions are collected between this time and the next
change of fraction mode or Off, where when the peak
detector encounters a peak, the peak is collected
independently of the time slices, specified by the
duration of the time-slices to collect in the
Time
slices
field)
• Peak-based with volume-slice recovery
(Volume-slice fractions are collected between this
time and the next change of fraction mode or Off,
where when the peak detector encounters a peak, the
peak is collected independently of the volume slices,
specified by the volume of the fractions to collect in
the
Volume slices
field)