7 Press  to start the adjustment. The CAL 
annunciator flashes to indicate the calibration is in 
progress. 
• Successful completion of each adjustment value is 
indicated by a short beep and the primary display 
briefly showing PASS.
• An adjustment failure is indicated by a long beep, the 
primary display showing FAiL and a calibration error 
number appearing in the secondary display. Check the 
input value, range, function, and entered adjustment 
value to correct the problem and repeat the adjustment 
step.
8 Repeat steps 3 through 7 for each gain adjustment point 
shown in the table.
9 Verify the AC Voltage Gain adjustments using the “AC 
Voltage Verif ication Test” on page 82.
Disp
Ta ble 1 7 AC Voltage Gain Adjustment
Input
Vrms Frequency Function
Frequency as 1kHz 
Cal Item 
10 mV 1kHz AC Voltage 10 mV
100 mV 1kHz 100 mV
1 V 1kHZ 1 V
10V 1kHz 10 V
100 V 1kHz 100 V
750 V 1kHz 750 V
Caution: Set the calibrator output to 0V before disconnecting from the multimeter input terminals.