136 34405A User’s and Service Guide
6 Specifications
Supplemental Measurement Specifications
Ta ble 2 8 Supplemental Measurement Specifications
DC Voltage
Measuring Method:
Sigma Delta A-to-D converter
Input Resistance:
10MΩ ± 2% range (typical)
Input Protection:
1000V on all ranges (HI terminal)
Resistance
Measurement Method:
2-wire Ohms
Open-circuit voltage:
Limited to < 5V
Input Protection:
1000V on all ranges (HI terminal)
DC Current
Shunt Resistance:
0.1Ω to 10Ω for 10mA to 1.2A ranges
0.01Ω for 12A range
Input Protection:
Front Panel 1.25A, 500V fuse for I terminal
Internal 15A, 600V fuse for 12A terminal
34405A users guide.book Page 136 Saturday, September 2, 2006 3:38 PM