98 34405A User’s and Service Guide
4 Performance Tests and Calibration
9 Verify the DC Voltage Gain adjustments using the “DC 
Voltage Gain Verif ication Test” on page 78.I
DC Current Gain Adjustment Procedure
Review the “Test Considerations” on page 71 and “Gain 
Adjustment Considerations” on page 95 sections before 
beginning this procedure. 
1 Press  to enter DC Current Gain Calibration. 
2 The primary display will show the uncalibrated value and 
the secondary display will show the reference value of Cal 
Item.  
3 Configure each Cal Item shown in the adjustment table 
below.
4 Use   (Auto)  or   (Range) to select the Cal 
Item.
Ta ble 1 5 DC Voltage Gain Adjustment
Input Function Cal Item  
Dual Banana Plug with copper wire short between 2 terminals DC Voltage Short 
100 mV 100 mV 
+ 1 V + 1 V 
- 1 V - 1 V 
10 V 10 V 
100 V 100 V 
1000 V 1000 V 
Caution: Set the calibrator output to 0V before disconnecting from the multimeter input terminals.
DCI
NOTE
If the zero adjustment procedure has been recently performed prior to the 
DC Current gain calibration procedure, the Cal Item 'Open' can be 
neglected.
34405A users guide.book  Page 98  Saturday, September 2, 2006  3:38 PM