140 34405A User’s and Service Guide
6 Specifications
General Characteristics
Ta ble 2 9 General Characteristics
Power Supply
• 100V/120V(127V)/220V(230V)/240V ± 10%
• AC line frequency 45Hz - 66Hz and (360Hz - 440Hz, 100/120V operation)
Power Consumption
• 16VA maximum, <11W average
Operating Environment
• Full accuracy at 0° C to 55° C
• Full accuracy to 80% RH at 30° C (non-condensing)
• Altitude up to 3000 meters
Storage Compliance
• - 40° C to 70° C
Safety Compliance
• Certified by CSA for IEC/EN/CSA/UL 61010-1 2
nd
Edition
Measurement Category
• CAT II, 300V: CAT I 1000Vdc, 750Vac rms, 2500Vpk transient over voltages
• Pollution degree 2
EMC Compliance
• Certified to IEC/EN 61326: 2002, CISPR 11, and equivalents for Group 1, Class A
Shock and Vibration
• Tested to IEC/EN 60086-2
Dimension (HxWxD)
• Rack: 88.5mm x 212.6mm x 272.3mm
• Bench: 103.8mm x 261.1mm x 303.2 mm
34405A users guide.book Page 140 Saturday, September 2, 2006 3:38 PM