EasyManuals Logo
Home>Agilent Technologies>Multimeter>34405A

Agilent Technologies 34405A User Manual

Agilent Technologies 34405A
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
Specifications 6
34405A User’s and Service Guide 143
Accuracy Specifications
Tran s fer Accur acy
Transfer accuracy refers to the error introduced by the
multimeter due to noise and short
term drift. This error
becomes apparent when comparing two nearly
equal signals
for the purpose of "transferring" the known accuracy of one
device to the other.
1Year Accuracy
These longterm accuracy specifications are valid at the
calibration temperature (T
cal
) ± 5 °C temperature range.
These specifications include the initial calibration errors plus
the multimeter's long
term drift errors.
Temperature Coefficients
Accuracy is usually specified at the calibration temperature
(T
cal
) ± 5 °C temperature range. This is a common
temperature range for many operating environments. You
must add additional temperature coefficient errors to the
accuracy specification if you are operating the multimeter
outside the ± 5 °C temperature range (the specification is
per °C).
34405A users guide.book Page 143 Saturday, September 2, 2006 3:38 PM

Table of Contents

Other manuals for Agilent Technologies 34405A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34405A and is the answer not in the manual?

Agilent Technologies 34405A Specifications

General IconGeneral
BrandAgilent Technologies
Model34405A
CategoryMultimeter
LanguageEnglish

Related product manuals