Performance Tests and Calibration 4
34405A User’s and Service Guide 109
3 Record the new Calibration Count (see page 109).
4 Press  and   simultaneously to exit the 
Adjustment Mode.  The instrument will be secured and 
returns to DC Voltage, autoranging measurements. 
Calibration Message
The instrument allows you to store a message in calibration 
memory. For example, you can store such information as the 
date when the last calibration was performed, the date when 
the next calibration is due, the instrument's serial number, 
or even the name and phone number of the person to 
contact for a new calibration. The calibration message may 
contain up to 40 characters.
You can record a calibration message only when the 
instrument is unsecured. You can read the calibration 
message (from remote only) whether the instrument is 
secured or unsecured.
To store a calibration message use the CALibration:STRing 
and CALibration:STRing? commands from the remote 
interface.
To Read the Calibration Count 
You can query the instrument to determine how many 
calibrations have been performed. Note that your instrument 
was calibrated before it left the factory. When you receive 
your instrument, read the count to determine its initial 
value. 
The count value increments by one for each calibration 
point, and a complete calibration will increase the value by 
many counts. The calibration count increments up to a 
maximum of  32767 after which it rolls over to 0. The 
calibration count can be read either remotely or from the 
Shift
DCV
34405A users guide.book  Page 109  Saturday, September 2, 2006  3:38 PM