68 34405A User’s and Service Guide
4 Performance Tests and Calibration
Calibration Overview
Closed - Case Electronic Calibration
The instruments features closed- case electronic calibration.
No internal mechanical adjustments are required. The
instrument calculates correction factors based upon the
input reference value you set. The new correction factors are
stored in non- volatile memory until the next calibration
adjustment is performed. Non - volatile EEPROM calibration
memory does not change when power has been off or after a
remote interface reset
Agilent Technologies Calibration Services
When your instrument is due for calibration, contact your
local Agilent Service Center for a low- cost recalibration. The
34405A is supported on automated calibration systems,
which allow Agilent to provide this service at competitive
prices.
Calibration Interval
A 1- year interval is adequate for most applications. Accuracy
specifications are warranted only if adjustment is made at
regular calibration intervals. Accuracy specifications are not
NOTE
Make sure you have read “Test Considerations” on page 71 before
calibrating the instrument.
34405A users guide.book Page 68 Saturday, September 2, 2006 3:38 PM

Table of Contents

Other manuals for Agilent Technologies 34405A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34405A and is the answer not in the manual?

Agilent Technologies 34405A Specifications

General IconGeneral
BrandAgilent Technologies
Model34405A
CategoryMultimeter
LanguageEnglish

Related product manuals