EasyManua.ls Logo

Agilent Technologies 34405A - To Calculate Total Measurement Error

Agilent Technologies 34405A
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
142 34405A User’s and Service Guide
6 Specifications
To Calculate Total Measurement Error
The multimeter's accuracy specifications are expressed in the
form:
( % of reading + % of range ). In addition to the reading
error and range error, you may need to add additional
errors for certain operating conditions. Check the list below
to make sure you include all measurement errors for a given
function. Also, make sure you apply the conditions as
described in the footnotes on the specification pages.
If you are operating the multimeter outside the
temperature range specified, apply an additional
temperature coefficient error.
For AC voltage and AC current measurements, you may
need to apply an additional low frequency error or crest
factor error.
34405A users guide.book Page 142 Saturday, September 2, 2006 3:38 PM

Table of Contents

Other manuals for Agilent Technologies 34405A

Related product manuals