EasyManua.ls Logo

Agilent Technologies 34405A - Page 105

Agilent Technologies 34405A
148 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Performance Tests and Calibration 4
34405A User’s and Service Guide 105
Frequency Gain Adjustment Procedure
Review the “Test Considerations on page 71 and “Gain
Adjustment Considerations” on page 95 sections before
beginning this procedure.
1 Press to enter Frequency Gain Calibration.
2 The primary display will show the calibration value and
the secondary display will show the reference value of Cal
Item.
3 Configure each Cal Item shown in the adjustment table
below.
4 Use (Auto) or (Range) to select the Cal
Item.
5 Apply the input signal shown in the Input Voltage and
Frequency column of the table.
Ta ble 1 9 Ohms Gain Adjustment
Input Function Cal Item
Dual Banana Plug with copper wire short between 2 terminals 2 - Wire Ohms Short
Input terminals open (remove any test leads or shorting plugs
from the input terminals)
Open
10 M Ω 10 MΩ
1 M Ω 1 MΩ
100 k Ω 100 k Ω
10 k Ω 10 k Ω
1 k Ω 1 k Ω
100 Ω 100 Ω
Freq
Freq
34405A users guide.book Page 105 Saturday, September 2, 2006 3:38 PM

Table of Contents

Other manuals for Agilent Technologies 34405A

Related product manuals