134 34410A/11A User’s Guide
5 Specifications
Triggering and Memory
Reading Hold Sensitivity: 1% of reading
Samples per Trigger: 1 to 50,000
Trigger Delay: 0 to 3600 sec (20 ms step size)
External Trigger Low–power TTL compatible input programmable edge triggered
Delay: < 1mS
Jitter: < 1mS
Max rate up to 5 kHz
Min Pulsewidth 1 us
Voltmeter Complete 3 V Logic output
Polarity Programmable edge pulse
Pulsewidth Approximately 2 ms
Non–volatile Memory 50,000 reading
Sample Timer
Range Up to 3600 sec in 20 ms steps
Jitter < 100 ns