34410A/11A User’s Guide 25
Quick Start 1
To Make a 2-Wire Resistance Measurement
Ranges: 100W, 1 kW, 10 kW, 100 kW, 1 MW, 10 MW, 100 MW
Maximum resolution: 100 mW (on 100W range)
Configurable parameters: INTEGRATION, RANGE, OFFSET COMP, AUTO ZERO, NULL,
and NULL VALUE
To nullout the test lead resistance:
1 Connect one end of the test leads at the meter, and short the probe
ends together.
2 Press null.
3 Connect the probe ends to the test circuit, and measure the corrected
resistance value.
To Make a 4-wire Resistance Measurement
Ranges: 100W, 1 kW, 10 kW, 100 kW, 1 MW, 10 MW, 100 MW
Maximum resolution: 100 mW (on 100W range)
Configurable parameters: INTEGRATION, RANGE, OFFSET COMP, AUTO ZERO, NULL,
and NULL VALUE