60 34410A/11A User’s Guide
2 Features and Functions
Remote Interface Operation: For all measurements except temperature,
diode check and continuity:
Autoranging may be enabled or disabled for the specified <function>
using this SENSe command:
SENSe:<function>:RANGe:AUTO {OFF>ONCE>0>ON>1};
where <function> = VOLTage:DC, VOLTage:AC, CURRent:DC,
CURRent:AC,RESistance, FRESistance or FREQency
The following command queries the autoranging feature status for the
specified <function>:
SENSe:<function>:RANGe:AUTO?
This query command returns “0” (OFF) or “1” (ON).
The following command manually sets the multimeter range for the
specified <function>:
SENSe:<function>:RANGe[:UPPER] {<range>>MIN>MAX>DEF}
Note: for <function> = FREQuency, this command changes the ac
input range.
A manual range may also be set using the CONFigure or MEASure
commands.
The following command queries the multimeter range setting for the
specified <function>:
SENSe:<function>:RANGe[:UPPER]? [{MIN>MAX}]
Refer to the Agilent 34410A/34411A Programmer’s Reference for
complete description and syntax for all commands on this page.

Table of Contents

Other manuals for Agilent Technologies 34410A

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 34410A and is the answer not in the manual?

Agilent Technologies 34410A Specifications

General IconGeneral
BrandAgilent Technologies
Model34410A
CategoryMultimeter
LanguageEnglish

Related product manuals