80 34410A/11A User’s Guide
2 Features and Functions
• After the programmed count has been reached, the display will show 
SAVING READINGS, then LOGGING COMPLETE. The readings are stored in 
non–volatile memory.
To stop a data logging event in progress, you may press either  , 
 or  . The primary display line will show DATA LOGGER? and 
offers selections of CONTINUE or STOP. The data logger continues to take 
readings while this is displayed. 
• If you select STOP and press  , the primary display will read DATA 
TO NV?, prompted you to SAVE or DISCARD the readings already logged. 
Select one of these options and press   to terminate the data 
logging function.
• If you select CONTINUE and press  , the data logging function will 
continue as programmed, until complete.
To read the data logged from the front panel, press   to display 
DATA LOGGER. Press   to select VIEW–RDGS, and press  . Press   or 
 to scroll through the stored readings.
• Remote Interface Operation: The following command returns all readings 
stored in non–volatile memory (NVMEM):
DATA:DATA? NVMEM
The following command deletes all readings stored in NVMEM.
DATA:DELete NVMEM
The following command returns the number of data points stored in 
NVMEM
DATA:POINts? NVMEM
Refer to the Agilent 34410A/34411A Programmer’s Reference for 
complete description and syntax for these commands.