EasyManua.ls Logo

Agilent Technologies 6490 - Page 19

Agilent Technologies 6490
112 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instrument Installation 1
Step 4. Connect the Triple Quadrupole LC/MS
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 19
Figure 10 Drain bottle
7 Install the gas filter(s):
For Triple Quadrupole LC/MSs without
AJS ESI source:
a Cut an appropriate length of the
1/4-inch PTFE gas tubing
(G1946-80078) and connect to the
nitrogen gas supply regulator or
nitrogen generator.
b Use the supplied nut and ferrule kit
(5183-0392) to attach the other
length of tubing to the RMSN-4 gas
filter. This configures the filter in
series with the supply tubing. See
Figure 11 as an example for
assembling the gas trap in the correct
configuration.
c Secure the gas filter in a vertical position. Use the supplied cable ties.
Spray chamber drain tube
Drain bottle vent
CDS inlet and LC waste fitting
Waste drain bottle
Secondary containment
(customer supplied)
Outlet to LC/MS
RMSN-4
Figure 11 Gas filter connected in
series
Inlet from nitrogen gas
source

Table of Contents

Other manuals for Agilent Technologies 6490

Related product manuals