3 Installation Completion and System Verification
Step 2. Reset the collision cell gas flow rate
56 Agilent 6400 Series Triple Quad LC/MS System Installation Guide
g Write this number down.
2 Edit the acquisition parameter file so that the newly determined value will
be applied whenever Data Acquisition starts up and will be used for tuning
and data acquisition:
a Open the D:\MassHunter\instr\default\ folder, and then open the
appropriate .xml file with Notepad:
b Locate the set point for the Collision Cell gas flow between the
<flowrate> and </flowrate> tags. In this example, the set point for the
flow rate is 520:
<cellFlow>
<flowA>1</flowA>
<flowB>0</flowB>
<flowrate>520</flowrate>
</cellFlow>
c Change the flow rate set point to the value that you wrote down as the
optimal flow rate. In this example, the value is changed to 550:
<cellFlow>
<flowA>1</flowA>
<flowB>0</flowB>
<flowrate>550</flowrate>
</cellFlow>
d Click File > Save to save the file.
e Click File > Exit to close the file.
3 Restart the Data Acquisition program:
a Close the Data Acquisition program.
6495 G6495AHWParams.xml
6490 G6490AHWParams.xml
6470 G6470AHWParams.xml
6460C G6460CAHWParams.xml
6460A G6460AHWParams.xml
6430A G6430AHWParams.xml
6420 G6420AHWParams.xml
6410 G6410BHWParams.xml

Table of Contents

Other manuals for Agilent Technologies 6490

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 6490 and is the answer not in the manual?

Agilent Technologies 6490 Specifications

General IconGeneral
BrandAgilent Technologies
Model6490
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals