Other Setup Tasks 5
To confirm the collision cell gas flow rate (6470A and 6495A/B only)
Agilent 6400 Series Triple Quad LC/MS System Installation Guide 97
Figure 47 Ramp parameter
g Click Ramp to ramp Prefilter DC 2 and find the DC value that
corresponds to the abundance maximum for m/z 290.
h Set the cursor to the maxima and click Accept.
Figure 48 Ramping Prefilter DC 2