General Maintenance 4
Troubleshooting and Maintenance Manual 117
Reinstallation
Materials needed
GC/MS interface heater assembly (G1099-60107)
Heater clamp (G1099-20210); replace the old clamp if it is damaged
Screwdriver, Torx T-15 (8710-1622)
Hex driver, 1.5 mm (8710-1570)
CAUTION
Installing a new heater and sensor in a damaged heater sleeve will result in poor
performance of the heated zone and could reduce the lifetime of the new parts.
Figure 16 Replacing the GC/MS interface heater and sensor
Heater clamp
Setscrew
Temperature sensor
Heater
Insulation
Cover
MS
Tip of interface