40 Troubleshooting and Maintenance Manual
2 General Troubleshooting
Temperature Symptoms
The MS has four heated zones:
Ion source
Front and rear mass filters
GC/MS interface
Each heated zone has a heater and temperature sensor. The ion source and
mass filters are powered and controlled by the MS. The GC/MS interface is
powered and controlled by the GC.
Ion source will not heat up
High-vacuum pump is off or has not reached normal operating conditions*
Incorrect temperature setpoint
Ion source has not had enough time to reach temperature setpoint
Ion source heater cartridge is not connected*
Ion source temperature sensor is not connected*
Ion source heater failed (burned out or shorted to ground)*
Ion source temperature sensor failed*
Source power cable is not connected to the quadrupole board*
MS electronics are not working correctly
* This will cause an error message.
Mass filter (quad) heaters will not heat up
High-vacuum pump is off or has not reached normal operating conditions*
Incorrect temperature setpoint
Mass filter has not had enough time to reach temperature setpoint
Mass filter heater cartridge is not connected*
Mass filter temperature sensor is not connected*
Mass filter heater failed (burned out or shorted to ground)*
Mass filter temperature sensor failed*
Cable is not connected to the quadrupole board*